Detail Ecotech Serinus 40 Oxides Of Nitrogen Analyzer
SERINUS 40
OXIDES OF NITROGEN
ANALYSER
The Serinus 40 Oxides of Nitrogen (NOX) Analyzer delivers
precise and reliable performance at excellent value. It uses
proven chemiluminescence technology to measure NO, NO2
and NOx in ambient air (LDL <0.4 ppb, range 0 - 20 ppm)
Benefits include:
• Reliable performance complemented with a two year warranty.
• Minimises time spent on site performing maintenance through superior remote instrument
control, diagnostic viewing and calibration.
• Comprehensive data logging and remote
viewing of over 200 operational parameters.
• Versatile interfacing through RS232, USB, Bluetooth, analog and digital I/O and optional TCP/IP
• Easy setup through an intuitive menu system, advanced GUI and a large alphanumeric keypad
with tactile keys.
• Instant status indication via illuminated instrument
status light buttons on the front panel.
• Removable flash memory stores ~10 years of data,
including up to 12 individual parameters and event
logs. Operational parameters can also be transferred
to memory for easy retrieval.
• Latest firmware updates can easily be installed using the USB flash memory drive.
• Programming, viewing, downloading and emailing of data is made even simpler using Bluetooth
connectivity combined with the “Ecotech Serinus Remote” Android App.
SPECIFICATIONS
Range: 0 - 20 ppm (autoranging)
Concentration units: mg/m3,ug/m3,ng/m3 ppm, ppb or ppt
Noise: < 0.2 ppb
Lower detectable limit: < 0.4 ppb or 0.5 % of reading whichever is greater
Linearity: ±1 % of full scale
Precision: 0.4 ppb or 0.5 % of reading whichever is greater
Zero drift (24 hours): < 0.5 ppb
Zero drift (7 days): < 0.5 ppb
Span drift (7 days): < 1 % of reading
Response time: 15 seconds to 95 %
STP reference: 0 °C, 20 °C, 25 °C at 101.3 kPa
Sample flow rate: 600 cc/min (total for 2 channels)
Temperature range: 0 - 40 °C
Operating power: 100 - 240 VAC, 50 - 60 Hz (autoranging)
Power consumption: 265 VA (max at start-up) 190 VA (nominal)
Dimensions: (W) 429 x (H) 175 x (D) 638 mm
Rack spacing: 3.5 RU
Weight: 21.9 kg
Communication:
• #1 RS232 port (supports all protocols*)
• #2 RS232 port (supports all protocols* and multidrop)
• USB port (Advanced protocol only)
• Bluetooth (Advanced protocol only)
• TCP/IP network port (optional, supports all protocols*)
Data logging:
8 GB removable USB flash memory drive that scores
the internal data logger, event log, automatic/manual
configuration backup, automatic/manual parameter list
save.
• Internal data logger (can log up to 12 of 200+ parameters)
• Logger interval (1 second to 24 hours)
• Storage capacity ~ 10 years (set to 1 minute interval)