Cooling Recirculating Chiller
Temperature control, incubation, material testing, corrosion testing, cell culture, tissue research, rotary evaporatororbital agitation at variable speeds to affect the growth of cell cultures.
Product Details
Temperature control, incubation, material testing, corrosion testing, cell culture, tissue research, rotary evaporatororbital agitation at variable speeds to affect the growth of cell cultures.




Features

● P.I.D temperature controller provides accurate and reliable temperature control.

● Large LCD display screen and interface provides for user-friendly operation.

● Preset On/Off function.

● Independent circulating pump switch for easy starting/stopping the cycle.

● Easy to set Adjustable Timer. (1 minute to 5,999 minutes)

● Maintenance-free operation with easy to clean surface.

● Both heater and bath chamber are made of corrosion-resistant stainless steel.

● No angle in bath chamber, easy to clean.

● Uniquely designed heating method can up to 80 °C .(BR-H series)

● Programming setting function with 7 periods and 9 steps for each period, which means. there are 63 programmable steps in total. (BR-H series)
Safety

●Temperature deviation alarm.

●Over current protection.

●Independent Over-temperature protection meets DIN 12880 International standard requirements.

●Liquid level display window, through which you can intuitively observe the liquid level in the tank.




Option

●Available RS-485 or USB ports for data collection.
Specification

Product Name
Cooling recirculating chiller
Model
BR-03
BR-05
BR-10
BR-20
BR-50
Storage Tank maximum capacity (L)
5
8
15.5
30
62
Temperature range

-20~20℃
Temperature accuracy
±2℃

Cooling capacity at 10°C (kw)
530
800
1500
2500
4000
Refrigerant
R404A

R410A

Security Features
Delay,leakage,overcurrent,overvoltage

Total power (W)
550
600
1300
1600
2400
Power requirements
AC220V/50HZ

Pump power
30
60

550
Pump flow max.(L/min)
6

17
30
Maximum head (m)
9

15

25
Inlet/Outlet pipe diameter (mm)
φ10
φ12
Weight
32

41
60
76
126
Liquid storage
3.5
5.2
13
25
52
Noise level
≤45
≤55



≤65



≤75
Dimensions W×D×H(mm) (include caster)
210×410×511
310×445×570
400×500×739
450×600×821
560×845×850