Back to Top

1
Hi, Guest!

  LOKASI :  Jakarta Timur

  AKTIVITAS :  11/05/2022

BAGIKAN :   

Bagikan :
Kontak Kami  
Jual Alat Lab Penguji Lingkungan 2. Distributor Alat Castle Inc.Indonesia 3.Jual alat General Laboratory 4.Distributor Alat Delta Ohm Indonesia.5.Distributor B
  • Gasboard
  • Gasboard-5260 Online Automobile Emission Gas Analyzer

Gasboard-5260 Online Automobile Emission Gas Analyzer

Update Terakhir
:
07 / 02 / 2022
Min. Pembelian
:
1 Unit

Harga

CALL
Bagikan
:
Dengan hormat, Kami PT.Mitra Katiga Sejahtera menjual alat laboratorium lingkungan.
I. STAPLEX, SHINAGAWA, NOVA GAS, DELTA OHM,LABTRON,TURNKEY INSTRUMENTS,HAZ DUST
E INSTRUMENT , EPAM-5000 ENVIRONMENTAL DEVICE , LABNICS,GOLD,AIRMETRIC,ASSEMBLAD,ANAPOL,ZHONGAN,KOMYO KITAGAWA,JXCT,TRAWAS
www.mitrakatigasejahtera.com

Detail Gasboard-5260 Online Automobile Emission Gas Analyzer

Gas Analyzer
Online Automobile Emission Gas Analyzer
Gasboard-5260 Automobile Emission Gas Analyzer adopts micro-flow NDIR technology with independent intellectual property and international PCT(CN2018100767) to measure NO gas, NDUV technology to measure NO2 gas and NDIR technology to measure HC, CO, CO2 gases simultaneously. Gasboard-5260 is widely used in gasoline vehicles emission gas analysis who requires direct measurements of NO&NO2 gases.

Features
Simultaneous measuring CO, CO2, HC, O2, NO, NO2 gases
Micro-flow NDIR NO gas bench replaces traditional ECD NO sensor, high accuracy, longer lifetime, no regular replacement
NDUV NO2 gas bench without NOx converter, direct NO2 detection with high accuracy
Auto temp. and pressure compensation
Multi-stage filtration system ensures longer lifetime

Specifications
Measurements

HC CO CO2 NO NO2 O2

Range

HC: (0~9999) ppm; CO: (0-15.00) % ; CO2: (0-20.00) % ; NO: (0~5000) ppm ; NO2: (0~1000) ppm ; O2: (0-25.00) %
Accuracy

HC: (0~2000) ppm: Abs. error: ±4ppm;Rel. error:±3%; (2001~5000) ppm:Rel. error:±5%; (5001~9999) ppm: Rel. error:±10%

CO: (0~10.00) %: Abs. error: ±0.02%;Rel. error:±3%; (10.01~15.00) %: Rel. error:±5%

CO2: (0~16.00) %: Abs. error:±0.3% VOL;Rel. error:±3%; (16.01~20.00) %: Rel. error:±5%

NO: (0~4000) ppm: Abs. error:±25ppm;Rel. error:±4%; (4001~5000) ppm: Rel. error:±8%

NO2: (0~1000) ppm:Abs. error:±25 ppm;Rel. error:±4%

O2: (0~25.00) %:Abs. error:±0.1%;Rel. error:±5%

* Remark:Abs. error or rel. error whichever is greater
Resolution

NO/NO2/HC:1ppm;CO2/CO/O2:0.01%;

Response time

HC/CO/CO2:T90≤8s; NO/NO2:T90≤10s; O2:T90≤12s

Warm up time

30min

Communication

RS-485/232