Back to Top

1
Hi, Guest!

  LOKASI :  Jakarta Timur

  AKTIVITAS :  11/05/2022

BAGIKAN :   

Bagikan :
Kontak Kami  
Jual Alat Lab Penguji Lingkungan 2. Distributor Alat Castle Inc.Indonesia 3.Jual alat General Laboratory 4.Distributor Alat Delta Ohm Indonesia.5.Distributor B
  • Gasboard
  • Gasboard-5230 Online Automobile Emission Gas Analyzer

Gasboard-5230 Online Automobile Emission Gas Analyzer

Update Terakhir
:
07 / 02 / 2022
Min. Pembelian
:
1 Unit

Harga

CALL
Bagikan
:
Dengan hormat, Kami PT.Mitra Katiga Sejahtera menjual alat laboratorium lingkungan.
I. STAPLEX, SHINAGAWA, NOVA GAS, DELTA OHM,LABTRON,TURNKEY INSTRUMENTS,HAZ DUST
E INSTRUMENT , EPAM-5000 ENVIRONMENTAL DEVICE , LABNICS,GOLD,AIRMETRIC,ASSEMBLAD,ANAPOL,ZHONGAN,KOMYO KITAGAWA,JXCT,TRAWAS
www.mitrakatigasejahtera.com

Detail Gasboard-5230 Online Automobile Emission Gas Analyzer

Gas Analyzer
Gasboard-5230
Online Automobile Emission Gas Analyzer
Gasboard-5230 Automobile Emission Gas Analyzer adopts micro flow NDIR principle with independent intellectual property and international PCT (CN2018100767) to detect NO, dual channel structure sensor and special filters highly resolve environmental temp. instability and electroniccomponents aging issues. It adopts NDIR principle to measure CO2 and NDUV principle to measure NO2 directly without NOx converter

Features
Micro-flow NDIR NO gas bench replaces traditional ECD NO sensor, high accuracy, longer lifetime, no regular replacement.
NDUV NO2 gas bench without NOx converter, direct NO2 detection with high accuracy.
Auto temp. and pressure compensation
Multi-stage filtration system ensures longer lifetime
Specifications
GasMeasurements

CO2 NO NO2

Range

CO2: (0.00-20.00) %; NO: (0~5000) ppm; NO2: (0~1000) ppm


Accuracy

CO2: (0.00~16.00) %:Abs. error:±0.3%;Rel. error:±3%; (16.01~20.00) %: Rel. error:±5%

NO: (0~4000) ppm:Abs. error:±25 ppm;Rel. error:±4%; (4001~5000) ppm:Rel. error:±8%

NO2: (0~1000) ppm:Abs. error:±25 ppm;Rel. error:±4%;

*Remark: Abs. error or Rel. error whichever is greater
Resolution

NO/NO2: 1ppm; CO2: 0.01%;

Response time

HC/CO/CO2: T90≤8s; NO/NO2: T90≤10s; O2: T90≤12s

Warm up time

30min

Communication

RS-485/232

Power supply

100~240VAC, 150w
Tampilkan Lebih Banyak